Teaching apparatus · Semiconductors and solid state
Hall effect as a function of temperature
New apparatus for the measurement of the Hall effect in a semiconductor (Ge) at various temperatures

The Hall effect is important in the study of material properties because it allows a direct measurement of the concentration of free charge carriers and of the sign of their charge, in both metals and semiconductors.
The Hall effect is essentially a manifestation of the Lorentz force F acting on every electric charge q moving with velocity v in a magnetic field B. In metals the electric current is carried only by electrons. In semiconductors, instead, the charge carriers can be both electrons and holes.
Consider a piece of conductor shaped as a parallelepiped (figure 1) immersed in a magnetic field B uniform and directed along the axis z, carrying an electric current of intensity Ix along x.

Figure 1: Geometry of the Hall effect: the Lorentz force FL on the moving charges acts in the direction shown by the arrow (regardless of the sign of the charge)
The electrons acquire a drift velocity vd = -μEx under the effect of the electric field Ex . The Lorentz force FL = q vd B, acts on the electrons in the direction of negative y and so they tend to accumulate on the face of the sample perpendicular to the axis y facing the viewer in figure 1. This accumulation of charges produces a potential difference VH between the two faces of the sample, and therefore an electric field EH (Hall field) such that the electric force qEH is equal and opposite to the Lorentz force: EH = vd B.
It is convenient to define the Hall coefficient as RH = EH/(JB), and it is easy to show that the following relation holds
RH = -1/(en) [or = +1/(ep)]
depending on whether it is a metal (or an N-doped semiconductor) or a P-doped semiconductor, where n and p are the densities of electrons and holes respectively.

Figure 2: Block diagram of the apparatus.
The semiconductor sample is placed in a copper box that can be inserted between the pole pieces of a magnet. A resistance heater wound on an aluminium cylinder at the base of the sample holder, and a cold finger that can be immersed in a liquid nitrogen bath inside the dewar, allow the temperature to be varied between 150 K and 450 K. The heater is switched on manually and is switched off automatically when the sample temperature exceeds 175 °C.
A thermocouple thermometer, through an AD8495 converter, records the temperature together with the values of the Hall voltage Vh and of the resistance R, through an interface connected to a Personal Computer.
7 wires are soldered to the sample in the configuration shown in figure 2.
Contacts 7 and 5 are used to measure the p.d. V=RI across the sample in order to obtain the electrical resistance R. Contacts 1 and 4 carry the current produced by the constant-current generator circuit.
The Hall voltage is taken between contact 6 and the wiper of potentiometer P, which is used to balance the signal for B=0.
A microprocessor in the control box calculates the sample temperature (with an interpolation polynomial of the e.m.f. values of the K thermocouple given by the NIST tables) and the resistance (from the measured value of the bias current).
The bias current and the gains of the two channels Vh and Vr can be selected from the front panel and all measured values are shown on an LCD display.

Figure 3: VH and R as a function of temperature
An example of the results of a measurement, on a sample biased with a current of 10 mA and immersed in a field of 0.4 T, is shown in figure 3.

Built-in gaussmeter
Semiconductors and solid state
See also
Request a quote
info@labtrek.it · +39 333 337 9944