To see how it works: click here (Video, 12.9 MB)
New version of the Haynes-Shockley experiment
The experiment proposed in 1949 by J.R. Haynes and W. Shockley to measure the drift mobility of electrons and holes in semiconductors is conceptually very simple, although it presents some difficulties in sample preparation, in the use of point contacts and in signal detection.
It is an experiment of great teaching value because it gives a direct view of the drift, diffusion and recombination of excess carriers.

Figure 1: Block diagram of the original Haynes and Shockley apparatus
Consider a bar of P-doped semiconductor, of length l, with two ohmic contacts soldered at its ends. Inside the crystal a sweeping electric field Esis produced by a pulsed voltage generator, shown in figure 1 as a battery plus a switch.
Two point contacts (electrodes E and C) are placed on the surface of the crystal, separated by a distance d .
The point contacts are (at least partially) rectifying and are therefore shown as diodes in figure 1.
If a negative pulse a few microseconds long, with an amplitude sufficient to forward-bias diode DE, is applied to electrode E (emitter), electrons are injected into the region of the crystal just below tip E.
This bunch of electrons, under the action of the electric field, starts moving to the right with drift velocity vd, and after a certain time t reaches the region of the crystal below tip C (collector).
The arrival of the bunch of electrons, by increasing the concentration of minority carriers in the region of contact C, causes an increase of the reverse current and therefore a drop in the potential at the end of the resistor connected to the collector.
On the screen of the oscilloscope, connected across resistor R, you first see a narrow negative pulse with an amplitude comparable to that of the injection pulse and, with a certain delay t (time of flight), a second negative pulse, wider and of smaller amplitude.
The first peak is simultaneous with the injection pulse: it is in fact the propagation signal of the electromagnetic field travelling in the crystal at the speed of light.
The second pulse instead corresponds to the passage of the bunch of electrons under the collector: its approximately Gaussian shape and its amplitude are determined by diffusion and recombination.
The collected pulse has a shape and an area that depend on the time of flight t, on the distance travelled d , on the diffusion coefficient D, and on the drift velocity: vd=μEs, where μ is the mobility of the electrons.
As they drift towards the collector, the injected electrons diffuse in all directions, so the detected pulse becomes wider and wider as the delay increases t.
The electrons also recombine with the holes they meet along the way, so the number of electrons that actually reach the collector decreases exponentially in time according to the law:
N(t) = No exp (-t/τ)
where τ is the lifetime of the charge carriers.
The version used here has an optical injection system for the excess carriers, which removes the difficulty of a good injecting contact.

Figure 2: HS with optical injection

Pulses in a P-doped sample for different values of the sweeping voltage
P-doped Germanium sample


I-V curve of the point contact Electron pulse for different fibre-collector distances
N-doped Germanium sample


I-V curve of the point contact Hole pulse for different fibre-collector distances
Ref.:An improved version of the Haynes-Shockley experiment (PDF, 0.2 MB) Am. J. Phys. 68, 80 (2000) (PDF, 0.2 MB)
Simulations (Java App) of the Haynes – Shockley signal:
Simulation 1. (Buffalo). Simulation 2. (Salamanca) https://electronica.usal.es/Haynes-Shockley.html in Spanish/English
Possible measurements for minority carriers (electrons in P-doped samples, and holes in N-doped samples) :
- Time of flight
- Drift velocity
- Mobility
- Lifetime
- Diffusion coefficient
The Haynes-Shockley apparatus includes:
- Sample holder with double slide for the optical fibre (motorized) and for the point contact (collector).
- Double pulser for the sweeping voltage and for the laser pulse
- Differential amplifier to subtract the sweeping voltage from the collected signal.
- Double polarity for P- or N-doped samples.
- Circuit to evaluate the rectifying behaviour of the point contact (I-V curves).
- LCD display shows the fibre-collector distance, sweeping voltage and laser pulse intensity.
- P-doped Germanium sample with ohmic contacts.
- Optional N-doped Germanium sample with ohmic contacts.
Videos and documents
- Haynes ENG.webm (Video, 12.9 MB)
- Haynes.pdf (PDF, 0.2 MB)